IPL / DPL · Hair removal & skin

LasePro IPL & DPL System

A dual-handpiece light platform. The HR handpiece is for hair removal and the SR handpiece for pigmentation, vascular concerns, acne and skin rejuvenation. Six preset programmes in Easy mode, full manual control in Expert mode.

  • Hair removal
  • Pigmentation
  • Vascular
  • Acne
  • Rejuvenation
  • Fine lines
Lazenix LasePro IPL and DPL machine with HR and SR handpieces
2Handpieces: HR and SR
6Preset programmes
I–VIFitzpatrick skin types
~500,000Shots per lamp

How it works

A xenon flash lamp produces broad-spectrum light. Filtering narrows it so that melanin in hair and pigment, or haemoglobin in vessels, absorbs the light and heats up.

DPL (Dynamic Pulsed Light) uses a narrower, more concentrated band of approximately 530–1200 nm. The HR handpiece is designed for hair removal with deeper penetration. The SR handpiece is designed for skin work: pigmentation, vessels, acne and texture.

Treatment courses

Hair removal
6–8 sessions. Face every 3–4 weeks, body every 4–6 weeks
Pigmentation
3–6 sessions, every 3–4 weeks
Vascular
3–5 sessions, every 3–4 weeks
Acne
4–8 sessions, every 1–2 weeks
Skin rejuvenation
4–6 sessions, every 3–4 weeks

Skin types

  • Fitzpatrick skin types I–VI
  • Reduced energy and extra caution for types V–VI
  • Patch test 24–48 hours before the first treatment

Two operating modes

Easy mode
Six preset programmes: hair removal, pigmentation, vascular, acne, skin rejuvenation, wrinkle
Expert mode
Manual control of energy, pulse width, pulse count, delay time and four pulse modes

Safety

  • Emergency stop button and key switch
  • Overheating and water-circulation warnings
  • Cooling: semiconductor, water and air
  • Protective eyewear for operator and client

Running costs

  • Flash lamp: approximately 500,000 shots
  • Distilled water, replaced monthly
  • Coupling gel

Full specification

Technology
IPL + DPL (Dynamic Pulsed Light)
Spectrum
530–1200 nm
Handpieces
HR (hair removal) + SR (skin rejuvenation), sapphire crystal tips
Energy density
1–50 J/cm², adjustable
Pulse
Independently adjustable multi-pulse
Preset programmes
6
Lamp life
Approximately 500,000 shots
Cooling
Semiconductor + water circulation + air
Display
12.1-inch HD capacitive touchscreen
Dimensions
50 × 59 × 126 cm
Weight
60 kg

Specifications are taken from the operator manual and may vary with configuration and software version. For professional use by trained practitioners.

Questions

What is the difference between the HR and SR handpieces?

The HR handpiece is designed for hair removal. The SR handpiece is designed for skin treatments such as pigmentation, vascular concerns, acne and skin rejuvenation.

How many IPL sessions does hair removal take?

The operator manual gives a course of 6–8 sessions, with the face treated every 3–4 weeks and the body every 4–6 weeks.

Which skin types can be treated with IPL?

LasePro IPL covers Fitzpatrick skin types I–VI, with reduced energy and extra caution for types V–VI. A patch test is carried out 24–48 hours before the first treatment.

Contact

Request a price for LasePro IPL

Tell us about your clinic. We'll reply with pricing, delivery details and answers to your questions.